616
Occupied Bandwidth Measurement
Meas Setup
Avg Mode
Enables you to set the averaging mode.
When set to Exponential (Exp) the measurement averaging continues using the specified number of
averages to compute each averaged value. The average is displayed at the end of each sweep.
When set to Repeat, the measurement resets the average counter each time the specified number of
averages is reached.
Max Hold (Remote Command Only)
When On, Max Hold displays and holds the maximum responses of the current measurement. Turn Max
Hold to Off to disable the maximum hold feature.
State Saved: Saved in instrument state.
Min: 1
Max: 10000
Backwards Compatibility SCPI: [:SENSe]:EBWidth:AVERage:COUNt
Initial S/W Revision: Prior to A.02.00
Modified at S/W Revision: A.03.00
Key Path: Meas Setup
Mode: WCDMA, C2K, WIMAX OFDMA, TD-SCDMA, 1xEVDO, LTE, LTETDD,
BLUETOOTH, WLAN
Remote Command: [:SENSe]:OBWidth:AVERage:TCONtrol EXPonential|REPeat
[:SENSe]:OBWidth:AVERage:TCONtrol?
Example: OBW:AVER:TCON REP
OBW:AVER:TCON?
Notes: You must be in the W-CDMA mode, cdma2000 mode, TD-SCDMA mode,
BLUETOOTH mode, LTE mode, LTE TDD mode, WLAN mode, 1xEVDO
mode, or WIMAX OFDMA mode to use this command.
Use:INSTrument:SELect to set the mode.
Preset: EXP
State Saved: Saved in instrument state.
Range: Exp | Repeat
Initial S/W Revision: Prior to A.02.00
Modified at S/W Revision: A.03.00
Key Path: SCPI Only