1380
Trigge r
Trig Slop eControls the trigger polarity. It is set positive to trigger on a rising edge and negative to trigger on a
falling edge.
Trig Delay
Controls a time delay during which the analyzer will wait to begin a sweep after meeting the trigger criteria. You can use negative delay to pre-trigger the instrument in time domain or FFT.
Example: TRIG:QINP:LEV 0.5 V
Preset: 0.25 V
State Saved: Saved in instrument state
Range: –1 to 1 V
Readback Text: 0.1 of displayed unit (V, mV, etc.)
Initial S/W Revision: Prior to A.02.00
Key Path: Trigger , Baseband I/Q , Inp ut Q
Remote Command: :TRIGger[:SEQuence]:QINPut:SLOPe POSitive | NEGative
:TRIGger[:SEQuence]:QINPut:SLOPe?
Example: TRIG:QINP:SLOP POS
Preset: POSitive
State Saved: Saved in instrument state
Initial S/W Revision: Prior to A.02.00
Key Path: Trigger , Baseband I/Q , Inp ut Q
Remote Command: :TRIGger[:SEQuence]:QINPut:DELay <time>
:TRIGger[:SEQuence]:QINPut:DELay?
:TRIGger[:SEQuence]:QINPut:DELay:STATe OFF|ON|0|1
:TRIGger[:SEQuence]:QINPut:DELay:STATe?
Example: TRIG:QINP:DEL 10 ms
TRIG:QINP:DEL:STAT ON
Preset: 1 us
OFF
State Saved: Saved in instrument state
Range: –2.5 s to +10 s
Initial S/W Revision: Prior to A.02.00