Spectrum Emission Mask Measurement
Meas Setup
Meas Setup
Displays the setup menu for the currently selected measurement.
Avg/Hold Num
Toggles averaging On or Off in addition to enabling you to set the number of measurement averages used to calculate the measurement result. The average is displayed at the end of each sweep. After the specified number of average counts, the average mode (termination control) setting determines the average action.In the remote mode, use the Average State command to turn averaging on or off.
Key Path: Front-panel key
Initial S/W Revision: Prior to A.02.00
Key Path: Meas Setup
Mode: WCDMA, C2K, WIMAX OFDMA, TD-SCDMA, 1xEVDO, LTE, LTETDD,
WLAN
Remote Command: [:SENSe]:SEMask:AVERage:COUNt <integer>
[:SENSe]:SEMask:AVERage:COUNt?
[:SENSe]:SEMask:AVERage[:STATe] ON|OFF|1|0
[:SENSe]:SEMask:AVERage[:STATe]?
Example: SEM:AVER:COUN 100
SEM:AVER:COUN?
SEM:AVER ON
SEM:AVER?
Notes: You must be in the mode that includes SEM measurement to use this
command. Use :INSTrument:SELect to set the mode.
Preset: 10
0FF
State Saved: Saved in instrument state.
Min: 1
Max: 10000
Initial S/W Revision: Prior to A.02.00
Modified at S/W Revision: A.02.00, A.03.00