241
System Functions
System
24 hours
With Auto Align Alert set to 24 Hours the instrument will signal an alert after a time span of 24 hours since the last
successful full alignment (for example, Align Now, All or completion of a full Auto Align). You may choose this
selection in an environment where the temperature is stable on a daily basis at a small risk of accuracy errors in
excess of the warranted specifications. The alert is the Error Condition message “Align Now, All required”.
For front-panel operation , confirmation is required to transition into this setting of Alert. The confirmation dialog
is:
No confirmation is required when Alert is configured through a remote command.
7 days
With Auto Align Alert is set to 7 days the instrument will signal an alert after a time span of 168 hours since the last
successful full alignment (for example, Align Now, All or completion of a full Auto Align). You may choose this
selection in an environment where the temperature is stable on a weekly basis, at a modest risk of accuracy
degradations in excess of warranted performance. The alert is the Error Condition message “Align Now, All
required”.
For front panel operation, confirmation is required for the customer to transition into this setting of Alert. The
confirmation dialog is:
Mode: All
Example: :CAL:AUTO:ALER TTEM
Readback Text: Time & Temp
Status Bits/OPC dependencies: Bit 14 is set in the Status Questionable Calibration register.
Initial S/W Revision: Prior to A.02.00
Key Path: System, Alignments, Auto Align, Alert
Mode: All
Example: :CAL:AUTO:ALER DAY
Readback Text: 24 hours
Status Bits/OPC dependencies: Bit 14 is set in the Status Questionable Calibration register.
Initial S/W Revision: Prior to A.02.00