436
ACP Measurement
Meas Setup
Limit Test

Turns limit checking for each offset On or Off. The limits may be specified within the Offset menu, for

each offset, both sides of the carrier. For results that fail the limit, a red F is appended. In the Combined

view, the bar turns red.

Example UNIT:ACP:POW:PSD DBMMHZ
UNIT:ACP:POW:PSD?
Couplings When the PSD unit is changed, the PSD reference result of the
“MEAS|READ|FETCH:ACP[n]?” is also changed by the PSD unit basis (in
either dBm/Hz or dBm/MHz).
Preset DBMHZ
State Saved Saved in instrument state.
Range dBm/Hz|dBm/MHz
Initial S/W Revision Prior to A.02.00
Modified at S/W Revision A.02.00, A.03.00
Key Path Meas Setup
Mode WCDMA, C2K, 1xEVDO, LTE
Remote Command :CALCulate:ACPower:LIMit:STATe OFF|ON|0|1
:CALCulate:ACPower:LIMit:STATe?
Example CALC:ACP:LIM:STAT OFF
CALC:ACP:LIM:STAT?
Notes You must be in the mode that includes ACP measurements to use this
command. Use :INSTrument:SELect to set the mode.
Preset WCDMA: ON
C2K: ON
1xEVDO: ON
LTE: ON
State Saved Saved in instrument state.
Range On|Off
Backwards Compatibility SCPI [:SENSe]:MCPower:LIMit[:STATe]
[:SENSe]:ACPower:LIMit[:STATe]
Initial S/W Revision Prior to A.02.00
Modified at S/W Revision A.02.00, A.03.00