Modulation Accuracy (Composite Rho)
AMPTD Y Scale
I/Q Error View – Phase Error Window
When Auto Scaling is On, pressing the Restart front-panel key will result in automatically displaying the scale per division and reference value results for the vertical axis in the phase error window.
Range On|Off
Key Path AMPTD Y Scale
Mode cdma2000
Remote Command :DISPlay:RHO:VIEW2:WINDow3:TRACe:Y[:SCALe]:COUPle
0|1|OFF|ON
:DISPlay:RHO:VIEW2:WINDow3:TRACe:Y[:SCALe]:COUPle?
Example DISP:RHO:VIEW2:WIND3:TRAC:Y:COUP ON
DISP:RHO:VIEW2:WIND3:TRAC:Y:COUP?
Restriction and Notes When this parameter is set to On, pressing the front-panel Restart key,
or the Restart key in the Meas Control menu, activates the scale
coupling function, that automatically determines scale per division and
reference values based on the measurement results. When you set a
value to either Scale/Div or Ref Value manually, Auto Scaling is
automatically set to Off.
Dependencies/Couplings See Restriction and Notes
Remote Command Notes WINDow3: Phase Error window on I/Q Error view
Preset ON
State Saved Saved in instrument state.
Range On|Off
Key Path AMPTD Y Scale