272
System Functions
System
Alert
Setting Alert to ON/OFF will enable/disable the display of RF Preselector alignment required message
on the status line. The instrument will power up with Alert On mode.
Schedule Setup
Enables you to schedule a task to run automatically at the background based on the recurrence and time
set in the scheduler. Make sure that the Instrument’s local time is accurate as the Scheduler relies on this
information to execute the task.
Task
There are 3 task that can be selected for the scheduler to run.
Task 1 is the 20 Hz to 30 MHz alignment
Task 2 is the 30 MHz to 3.6 GHz alignment
Task 3 is the 20 Hz to 3.6 GHz alignment.
Key Path: System, Alignments, RF Preselector
Mode: All
Remote Command: :CALibration:RFPSelector:ALERt ON|OFF|0|1
:CALibration:RFPSelector:ALERt?
Example: :CAL:RFPS:ALER OFF
Notes: For model N9038A only.
Error Condition will be generated when the alert is On and any of the RF
Preselector alignments has expired.
Preset: This is unaffected by Preset, but is set to ON on a “Restore System
Defaults->Align”.
State Saved: No
Initial S/W Revision: A.08.00
Key Path: System, Alignments, RF Preselector
Initial S/W Revision: A.08.00
Key Path: System, Alignments, RF Preselector, Schedule Setup
Mode: All
Remote Command: :CALibration:RFPSelector:SCHeduler:TASK T1|T2|T3
:CALibration:RFPSelector:SCHeduler:TASK?
Example: :CAL:RFPS:SCH:TASK T1