1064
Meas Setup
Frequency is 12.5% above the range between the minimum and maximum Frequency so that span
exceeds this range by one graticule division on either side.
For time limits and a time-domain X-axis, sets the sweep time to match the maximum Time of the
selected Limit.
If the domain of the selected limit does not match the domain of the X-axis, no action is taken. Standard
clipping rules apply, if the value in the table is outside the allowable range for the X axis.
Delete Limit
Deletes the currently selected limit line. Pressing Delete Limit purges the data from the limit line tables.
Limit data – including secondary parameters such as description, margin value, et cetera - are cleared
and returned to factory preset settings.
When this key is pressed a prompt is placed on the screen that says “Please press Enter or OK key to
delete limit. Press ESC or Cancel to close this dialog.” The deletion is only performed if you press OK or
Enter; if so, after the deletion, the informational message “Limit deleted” appears in the MSG line.
Test Lim its
Selects whether displayed traces are tested against displayed limits (i.e. those for which Limit On/Off is
set to On).
For each displayed trace for which a Limit is turned on, a message is displayed in the upper-left corner of
the graticule to notify whether the trace passes or fails the limits.
If the trace is at or within the bounds of all applicable limits and margins, the text “Trace x Pass” is
displayed in green, where x is the trace number. A separate line is used for each reported trace.
If the trace is at or within the bounds of all applicable limits, but outside the bounds of some applicable
margin, the text “Trace x Fail Margin” is displayed in amber, where x is the trace number. A separate line
is used for each reported trace.
If the trace is outside the bounds of some applicable limits, the text “Trace x Fail” is displayed in red,
where x is the trace number. A separate line is used for each reported trace.
If the trace has no enabled limits, or the trace itself is not displayed, no message is displayed for that
trace.
The PASS/FAIL box in the corner of the Meas Bar is only displayed if there is at least one “Trace
Pass/Fail” indication displayed in the graticule.
Key Path Meas Setup, Limits, Edit
Initial S/W Revision A.02.00
Key Path Meas Setup, Limits
Remote Command :CALCulate:LLINe[1]|2|3|4|5|6:DELete
Example :CALC:LLIN2:DEL deletes all data for limit line 2.
Initial S/W Revision A.02.00