Spectrum Emission Mask Measurement
Meas Setup
Sweep TimeSpecifies the sweep time for the currently selected offset and enables you to toggle this function On or Off for each offset.
State Saved: Saved in instrument state.
Min: Start Freq plus (+) 100 Hz (for that offset)
Max: 500 MHz
Initial S/W Revision: Prior to A.02.00
Modified at S/W Revision: A.02.00, A.03.00, A.10.00, A.11.00
Key Path: Meas Setup, Offset/Limits
Mode: WCDMA, C2K, WIMAX OFDMA, TD-SCDMA, 1xEVDO, LTE, LTETDD,
WLAN
Remote Command: [:SENSe]:SEMask:OFFSet[1]|2:LIST:SWEep:TIME <time>, …
[:SENSe]:SEMask:OFFSet[1]|2:LIST:SWEep:TIME?
[:SENSe]:SEMask:OFFSet[1]|2:LIST:SWEep:TIME:AUTO
ON|OFF|1|0, …
[:SENSe]:SEMask:OFFSet[1]|2:LIST:SWEep:TIME:AUTO?
Example: SEM:OFFS2:LIST:SWE:TIME 1.0 ms, 3.4 ms, 2.08 ms, 1.0 ms, 1.0 ms, 1.0
ms
SEM:OFFS2:LIST:SWE:TIME?
SEM:OFFS2:LIST:SWE:TIME:AUTO ON, ON, ON, ON, OFF, OFF
SEM:OFFS2:LIST:SWE:TIME:AUTO?
Notes: Comma separated list of values.
OFFSet1 is for BTS, 2 for MS. Default is BTS.
You must be in the mode that includes SEM measurement to use this
command. Use :INSTrument:SELect to set the mode.
Couplings: When the sweep time is set manually, Mode coupling is set to MANual
Preset: Automatically calculated
Modes (except WLAN): ON,ON,ON,ON,ON,ON
WLAN: ON,ON,ON,ON,ON,ON,ON,ON, ON,ON,ON,ON
State Saved: Saved in instrument state.
Min: 1 ms
Max: 10 s
Backwards Compatibility SCPI: [:SENSe]:SEMask:OFFSet[1]|2:LIST:SWEep[:TIME]
Initial S/W Revision: Prior to A.02.00