1382
Trigge r
Trig Delay
Controls a time delay during which the analyzer will wait to begin a sweep after meeting the trigger
criteria. You can use negative delay to pre-trigger the instrument in time domain or FFT.
Trigger Center Frequency
This key sets the center frequency to be used by the auxiliary receiver.
Trigger Bandwidth
This key sets the information bandwidth used by the auxiliary receiver for the Auxiliary Channel I/Q
Magnitude trigger.
Key Path: Trigger, Baseband I/Q, Aux Channel I/Q Mag
Remote Command: :TRIGger[:SEQuence]:AIQMag:DELay <time>
:TRIGger[:SEQuence]:AIQMag:DELay?
:TRIGger[:SEQuence]:AIQMag:DELay:STATe OFF|ON|0|1
:TRIGger[:SEQuence]:AIQMag:DELay:STATe?
Example: TRIG:AIQM:DEL 10 ms
TRIG:AIQM:DEL:STAT ON
Preset: 1 us
OFF
State Saved: Saved in instrument state
Range: –2.5 s to +10 s
Initial S/W Revision: Prior to A.02.00
Key Path: Trigger, Baseband I/Q, Aux Channel I/Q Mag
Remote Command: :TRIGger[:SEQuence]:AIQMag:CENTer <freq>
:TRIGger[:SEQuence]:AIQMag:CENTer?
Example: :TRIG:AIQM:CENT 10 MHz
Notes: Trigger CF + 1/2 Trigger BW < Max
Trigger CF – 1/2 Trigger BW > Min
Preset: 0 Hz
State Saved: Saved in instrument state
Range: –40 MHz to 40 MHz
Initial S/W Revision: Prior to A.02.00
Key Path: Trigger, Baseband I/Q, Aux Channel I/Q Mag