1374
Trigge r
line will be displayed to indicate the trigger level.
Trig Slop e
Controls the trigger polarity. It is set positive to trigger on a rising edge and negative to trigger on a
falling edge.
Trig Delay
Controls a time delay during which the analyzer will wait to begin a sweep after meeting the trigger
criteria. You can use negative delay to pre-trigger the instrument in time domain or FFT.
Key Path: Trigger, Baseband I/Q, I/Q Mag
Remote Command: :TRIGger[:SEQuence]:IQMag:LEVel <ampl >
:TRIGger[:SEQuence]:IQMag:LEVel?
Example: TRIG:IQM:LEV –30 dBm
Notes: The I/Q reference impedance is used for converting between power and
voltage.
Preset: –25 dBm
State Saved: Saved in instrument state
Range: –200 dBm to 100 dBm
Readback Text: <level> dBm
Initial S/W Revision: Prior to A.02.00
Key Path: Trigger, Baseband I/Q, I/Q Mag
Remote Command: :TRIGger[:SEQuence]:IQMag:SLOPe POSitive | NEGative
:TRIGger[:SEQuence]:IQMag:SLOPe?
Example: TRIG:IQM:SLOP POS
Preset: POSitive
State Saved: Saved in instrument state
Initial S/W Revision: Prior to A.02.00
Key Path: Trigger, Baseband I/Q, I/Q Mag
Remote Command: :TRIGger[:SEQuence]:IQMag:DELay <time>
:TRIGger[:SEQuence]:IQMag:DELay?
:TRIGger[:SEQuence]:IQMag:DELay:STATe OFF|ON|0|1
:TRIGger[:SEQuence]:IQMag:DELay:STATe?