742
Modulation Accuracy (Composite Rho)
Marker
Couple Markers
Toggles the state of the markers to be coupled On or Off. When this function is true (On), moving any
marker causes an “equal X Axis movement” of every other marker which is active. By “equal X Axis
movement” we mean that we preserve the difference between each marker’s X Axis value (in the
fundamental x-axis units of the trace that marker is on) and the X Axis value of the marker being moved
(in the same fundamental x-axis units).
See Couple Markers in the "Marker" section for more information.
This may result in markers going off screen.
All Markers Off
Turns off all markers.
Preset EVM
State Saved Saved in instrument state.
Range POLar |EVM|MERRor|PERRor|
Key Path Marker
Mode cdma2000
Remote Command :CALCulate:RHO:MARKer:COUPle[:STATe] ON|OFF|1|0
:CALCulate:RHO:MARKer:COUPle[:STATe]?
Example CALC:RHO:MARK:COUP ON
CALC:RHO:MARK:COUP?
Restriction and Notes When the marker is assigned to the IQ Measured Polar graph, a Chip
value is coupled instead of an X Axis value.
Preset OFF
State Saved Saved in instrument state.
Range On|Off
Key Path Marker
Mode cdma2000
Remote Command :CALCulate:RHO:MARKer:AOFF
Example CALC:RHO:MARK:AOFF
Key Path Marker