423
ACP Measurement
Meas Setup
Filter BW
Selects a Gaussian filter based on its –3 dB (Normal) bandwidth or its –6 dB bandwidth.
LimitsLimits key accesses a menu of functions that contains Select Offset, Abs Limit, Rel Limit and Fail Mask
Notes Offset sub op code. 1 for BTS, 2 for MS. Default is BTS.
You must be in the mode that includes ACP measurements to use this
command. Use :INSTrument:SELect to set the mode.
Couplings See the description above
Preset GAUSsian, GAUSsian, GAUSsian, GAUSsian, GAUSsian, GAUSsian
State Saved Saved in instrument state.
Range GAUSsian|FLATtop
Backwards Compatibility SCPI [:SENSe]:ACPower:OFFSet[1]|2:LIST:BWIDth:SHAPe
Initial S/W Revision Prior to A.02.00
Modified at S/W Revision A.02.00, A.03.00
Key Path Meas Setup, Offset/Limits, Offset BW, RBW Control
Mode WCDMA, C2K, WIMAX OFDMA, TD-SCDMA, 1xEVDO, LTE, LTETDD
Remote Command [:SENSe]:ACPower:OFFSet[1]|2:LIST:BANDwidth:TYPE
DB3|DB6, DB3|DB6, DB3|DB6, DB3|DB6, DB3|DB6, DB3|DB6
[:SENSe]:ACPower:OFFSet[1]|2:LIST:BANDwidth:TYPE?
Example ACP:OFFS2:LIST:BAND:TYPE DB3,DB3,DB3,DB3,DB3,DB3
ACP:OFFS2:LIST:BAND:TYPE?
Notes Offset sub op code. 1 for BTS, 2 for MS. Default is BTS.
You must be in the mode that includes ACP measurements to use this
command. Use :INSTrument:SELect to set the mode.
Dependencies Grayed out unless the Gaussian filter type is selected
Preset DB3, DB3, DB3, DB3, DB3, DB3
State Saved Saved in instrument state.
Range –3 dB (Normal)| –6 dB
Backwards Compatibility SCPI [:SENSe]:ACPower:OFFSet[1]|2:LIST:BWIDth:TYPE
Initial S/W Revision Prior to A.02.00
Modified at S/W Revision A.02.00, A.03.00