733
Modulation Accuracy (Composite Rho)
AMPTD Y Scale
I/Q Error View – Phase Error Window
Sets the reference position of the vertical axis in the phase error window.
Auto Scaling
Toggles the Auto Scaling function between On and Off. When Auto Scaling is On, pressing the Restart
front-panel key results in automatically determining scale per division and reference values based on the
measurement results.
When you set a value to either Scale/Div or Ref Value manually, Auto Scaling automatically changes to
Off.
I/Q Error View - EVM Window
When Auto Scaling is On, pressing the Restart front-panel key results in automatically displaying the
scale per division and reference value results for the vertical axis in the EVM window.
Example DISP:RHO:VIEW2:WIND2:TRAC:Y:RPOS TOP
DISP:RHO:VIEW2:WIND2:TRAC:Y:RPOS?
Preset CENTer
State Saved Saved in instrument state.
Range Top|Ctr|Bot
Key Path AMPTD Y Scale
Mode cdma2000
Remote Command :DISPlay:RHO:VIEW2:WINDow3:TRACe:Y[:SCALe]:RPOSiti
on TOP|CENTer|BOTTom
:DISPlay:RHO:VIEW2:WINDow3:TRACe:Y[:SCALe]:RPOSiti
on?
Example DISP:RHO:VIEW2:WIND3:TRAC:Y:RPOS TOP
DISP:RHO:VIEW2:WIND3:TRAC:Y:RPOS?
Preset CENTer
State Saved Saved in instrument state.
Range Top|Ctr|Bot
Key Path AMPTD Y Scale
Key Path AMPTD Y Scale
Mode cdma2000