300
System Functions
System
Show RFIO Self Test Results contents (Remote Command Only)
A remote command is available to obtain the contents of the Show RFIO Self Test Results screen (the entire
contents, not just the currently displayed page).
FEC Self Test REsults
Provides a display of last FEC test results, the display should appear listing model number, serial number and test
time at the top of display, and then list test date/time, test name, measured value, valid range and pass/fail of each
FEC test item, the tabular data should be directly printable.
This function is NOT available on EXT E6607A and E6607B models.
11/23/2012 16:13:57 ABUS_DET_SYNTH 29.879 14.400 - 100.000 Pass
11/23/2012 16:13:57 ABUS_DET_LO 18.136 9.000 - 100.000 Pass
11/23/2012 16:13:57 ABUS_DET_MOD 13.556 6.600 - 100.000 Pass
11/23/2012 16:13:57 ABUS_DET_MOD_FLT 18.000 7.800 - 100.000 Pass
FilterBankTest1
Date Time(GMT) Name MeasValue ValidRange Result
11/23/2012 16:13:57 Ban k1_LPF_ 550MHz 30.357 >= 10.000 Pass
11/23/2012 16:13:57 Ban k1_LPF_ 750MHz 29.358 >= 10.000 Pass
11/23/2012 16:13:57 Bank1_ LPF_1020MHz 27.036 >= 10.000 Pass
11/23/2012 16:13:57 Bank1_ LPF_1600MHz 27.594 >= 10.000 Pass
11/23/2012 16:13:57 Bank1_ LPF_2400MHz 21.490 >= 10.000 Pass
11/23/2012 16:13:57 Bank1_ LPF_3000MHz 18.476 >= 10.000 Pass
11/23/2012 16:13:57 OFFSET_QP 0.121 0.100 - 0.125 Pass
Remote Command: :SYSTem:TEST:WCTS:MPADapter:RESult?
Example: SYST:TEST:WCTS:MPAD:RES?
Notes: The output is an IEEE Block format of the Show RFIO Self Test Results
contents. Each line is separated with a new-line character.
Initial S/W Revision: A.12.50
Key Path: System, Diagnostics, Self Test,Show Results
Remote Command: :SYSTem:TEST:WCTS:SHOW:RESult FEC
Example: SYST:TEST:WCTS:SHOW:RES FEC
Initial S/W Revision: A.12.50