346
Channel Power Measurement
Meas Setup
PSD Unit

Sets the unit bandwidth for Power Spectral Density. The available units are dBm/Hz and dBm/MHz.

Meas Preset
Restores all the measurement parameters to their default values.
Key Path: Meas Setup
Mode: WCDMA, C2K, WIMAX OFDMA, 1xEVDO, LTE, LTETDD, WLAN
Remote Command: :UNIT:CHPower:POWer:PSD DBMHZ|DBMMHZ
:UNIT:CHPower:POWer:PSD?
Example: UNIT:CHP:POW:PSD DBMMHZ
UNIT:CHP:POW:PSD?
Couplings: When the PSD unit is changed, the PSD result of the
“MEAS|READ|FETCH:CHP1?” is also changed by the PSD unit basis (in
either dBm/Hz or dBm/MHz).
Preset: DBMHZ
WLAN: DBMMHZ
State Saved: Saved in instrument state.
Range: dBm/Hz|dBm/MHz
Initial S/W Revision: Prior to A.02.00
Modified at S/W Revision: A.02.00, A.03.00
Key Path: Meas Setup
Mode: WCDMA, C2K, WIMAX OFDMA, 1xEVDO, LTE, LTETDD, WLAN
Remote Command: :CONFigure:CHPower
Example: CONF:CHP
Initial S/W Revision: Prior to A.02.00
Modified at S/W Revision: A.02.00, A.03.00