1295
Sweep/Control
command.
Gate Polarity (Remote Command Only)
Sets the polarity for the gate signal. This setup is now done using the gate trigger’s slope setting.
When Positive (Pos) is selected, a positive-going edge (Edge) or a high voltage (Level) will satisfy the
gate condition, after the delay set with the Gate Delay key. When Negative (Neg) is selected, a
negative-going edge (Edge) or a low voltage (Level) will satisfy the gate condition after the delay.
Points
Sets the number of points taken per sweep, and displayed in the traces. The current value of points is
displayed parenthetically, next to the sweep time in the lower-right corner of the display. Using more
points provides greater resolution. Using fewer points compacts the data and decreases the time required
to access a trace over the remote interface.
Increasing the number of points does not increase the sweep time; however, it can slightly impact the
trace processing time and therefore the overall measurement speed. Decreasing the number of points
Remote Command: [:SENSe]:SWEep:EGATe:EXTernal[1]|2:LEVel <voltage>
[:SENSe]:SWEep:EGATe:EXTernal[1]|2:LEVel?
Notes: This command is simply an alias to
:TRIGger[:SEQuence]:EXTernal[1]|2:LEVel
For details refer
Initial S/W Revision: Prior to A.02.00
Remote Command: [:SENSe]:SWEep:EGATe:POLarity NEGative|POSitive
[:SENSe]:SWEep:EGATe:POLarity?
Example: SWE:EGAT:POL NEG
SWE:EGAT:POL?
Preset: POSitive
State Saved: Saved in instrument state
Backwards Compatibility SCPI: [:SENSe]:SWEep:TIME:GATE:POLarity ESA compatibility
Initial S/W Revision: Prior to A.02.00
Remote Command: [:SENSe]:SWEep:TIME:GATE:LEVel HIGH|LOW
[:SENSe]:SWEep:TIME:GATE:LEVel?
ESA compatibility
Preset: HIGH
Initial S/W Revision: Prior to A.02.00