729
Modulation Accuracy (Composite Rho)
AMPTD Y Scale
I/Q Error View – Mag Error Window
Sets the absolute power reference value in the magnitude error window.
I/Q Error View – Phase Error Window
Sets the absolute power reference value in the phase error window.
Mode cdma2000
Remote Command :DISPlay:RHO:VIEW2:WINDow2:TRACe:Y[:SCALe]:RLEVel
<real>
:DISPlay:RHO:VIEW2:WINDow2:TRACe:Y[:SCALe]:RLEVel?
Example DISP:RHO:VIEW2:WIND2:TRAC:Y:RLEV 0.0
DISP:RHO:VIEW2:WIND2:TRAC:Y:RLEV?
Restriction and Notes If Auto Scaling is set to On, this value is automatically determined by
the measurement result. When you set this value manually, Auto
Scaling is automatically set to Off.
Dependencies/Couplings See Restriction and Notes
Remote Command Notes WINDow2: Mag Error window on I/Q Error view
Preset 0.00
State Saved Saved in instrument state.
Min –500.0
Max 500.0
Key Path AMPTD Y Scale
Mode cdma2000
Remote Command :DISPlay:RHO:VIEW2:WINDow3:TRACe:Y[:SCALe]:RLEVel
<real>
:DISPlay:RHO:VIEW2:WINDow3:TRACe:Y[:SCALe]:RLEVel?
Example DISP:RHO:VIEW2:WIND3:TRAC:Y:RLEV 0.0
DISP:RHO:VIEW2:WIND3:TRAC:Y:RLEV?
Restriction and Notes If Auto Scaling is set to On, this value is automatically determined by
the measurement result. When you set this value manually, Auto
Scaling is automatically set to Off.
Dependencies/Couplings See Restriction and Notes
Remote Command Notes WINDow3: Phase Error window on I/Q Error view
Preset 0.00
State Saved Saved in instrument state.