640
Occupied Bandwidth Measurement
Trace/Detector
Trace/Detector
Accesses a menu of functions that enable you to control the detectors for the current measurement.
Trac e Typ e
Allows you to select the type of trace you want to you use for the current measurement. The first page of this menu contains a 1–of–N selection of the trace type (Clear Write, Average, Max Hold, Min Hold) for the selected trace.
Key Path: Front-panel key
Initial S/W Revision: Prior to A.02.00
Key Path: Trace/Detector
Mode: WCDMA, C2K, WIMAX OFDMA, TD-SCDMA, 1xEVDO, LTE, LTETDD,
BLUETOOTH, WLAN
Remote Command: :TRACe:OBWidth:TYPE WRITe|AVERage|MAXHold|MINHold
:TRACe:OBWidth:TYPE?
Example: TRAC:OBW:TYPE MINH
TRAC:OBW:TYPE?
Notes: WRITe = Clear Write
AVERage = Average
MAXHold = Maximum Hold
MINHold = Minimum Hold
Couplings: When Detector setting is “Auto” ([:SENSe]:OBWidth:DETector:AUTO?),
Detector ([:SENSe]:OBWidth:DETector[:FUNCtion]?) switches aligning
with the switch of this parameter: “NORMal” with WRITe (Clear Write),
“AVERage” with AVERage, “POSitive (peak)” with MAXHold, and
“NEGative (peak)” with MINHold.
Preset: AVERage
BLUETOOTH: MAX HOLD.
State Saved: Saved in instrument state.
Range: WRITe|AVERage|MAXHold|MINHold
Initial S/W Revision: Prior to A.02.00
Modified at S/W Revision: A.03.00