1184
Source (Internal)
Buffered Trigger
Selects Buffered Trigger as the trigger response for single trigger type. Buffered Trigger sets the waveform
generator to play a waveform sequence or segment once when a trigger is received. If a trigger is received during
playback, the waveform generator plays the sequence or segment to the end, then plays the sequence or segment
once more.
Restart on Trigger
Selects Restart on Trigger as the trigger response for single trigger type. Restart on Trigger sets the waveform
generator to play a waveform sequence or segment once when a trigger is received. If a trigger is received during
playback, the waveform generator resets and plays the sequence or segment from the start.
Segment Advance
Sets the active trigger type to Segment Advance. If Segment Advance is already selected as the active trigger type,
pressing this key allows access to the segment advance trigger type setup menu.
Segment Advance triggering allows you to control the playback of waveform segments within a waveform
sequence. When a trigger is received the ARB advances to the next waveform segment within the waveform
sequence. This type of triggering ignores the repetition count for the waveform segment within the waveform
sequence. For example, if a waveform segment has a repetition count of 10 and you select single segment advance
triggering mode, the waveform segment will only play once.
Segment Advance triggering can also be used for waveform segments only. In this situation the same waveform
segment is played again when a trigger is received.
Single
Selects Single as the trigger response for Segment Advance trigger type. With single selected, once a trigger is
received a segment is played once. If a trigger is received during playback of a segment, the segment plays to
Key Path: Source, Modulation Setup, ARB, Trigger Type, Single
Example: SOUR:RAD:ARB:RETR ON
Initial S/W Revision: A.05.00
Key Path: Source, Modulation Setup, ARB, Trigger Type, Single
Example: SOUR:RAD:ARB:RETR IMM
Initial S/W Revision: A.05.00
Key Path: Source, Modulation Setup, ARB, Trigger Type
Remote Command: :SOURce:RADio:ARB:TRIGger:TYPE:SADVance[:TYPE]
SINGle|CONTinuous
:SOURce:RADio:ARB:TRIGger:TYPE:SADVance[:TYPE]?
Example: SOUR:RAD:ARB:TRIG:TYPE:SADV SING
Preset: CONTinuous
Range: Single | Continuous
Initial S/W Revision: A.05.00