Modulation Accuracy (Composite Rho)
SPAN X Scale
SPAN X Scale
Accesses a menu of functions that enable you set the horizontal scale parameters.
This menu is available when in the “I/Q Error” view. It is not available when in the following views:
“I/Q Measured Polar Graph” or “Power, Timing and Phase”
Ref Value
Sets the chip reference value on the horizontal axis. The default setting is 0.000 chips. When Auto
Scaling is set to On, the displayed graphs use a Scale/Div value determined by the test set, based on the
measurement result.
I/Q Error View - EVM Window
Sets the chip reference value on the horizontal axis in the EVM window.
Key Path Front panel key
Key Path Span X Scale
Mode cdma2000
Remote Command :DISPlay:RHO:VIEW2:WINDow[1]:TRACe:X[:SCALe]:RLEVe
l <real>
:DISPlay:RHO:VIEW2:WINDow[1]:TRACe:X[:SCALe]:RLEVe
l?
Example DISP:RHO:VIEW2:WIND:TRAC:X:RLEV 0.0
DISP:RHO:VIEW2:WIND:TRAC:X:RLEV?
Restriction and Notes If Auto Scaling is set to On, this value is automatically determined by
the measurement result. When you set this value manually, Auto
Scaling is automatically set to Off.
Dependencies/Couplings See Restriction and Notes
Remote Command Notes WINDow1: Evm Error window on I/Q Error view
Preset 0.0
State Saved Saved in instrument state.
Min 0.0
Max 5000000.0
Key Path Span X Scale