576
Spectrum Emission Mask Measurement
View/Display
Trace Window
Results Window
Rel Pwr FreqSets the display to the Relative Peak Power and Frequency view. The views differ depending on the
setting of the measurement type (Meas Type) under the Measurement Setup menu.
“Rel Peak Pwr & Freq (Total Pwr Ref)” on page 576
“Rel Peak Pwr & Freq (PSD Ref)” on page 579
“Rel Peak Pwr & Freq (Spectrum Pk Ref)” on page 581
Rel Peak Pwr & Freq (Total Pwr Ref)
This view consists of the following two windows:
Corresponding Trace yellow - Combined trace from carrier and each offset
Name Corresponding Results
Total Pwr Absolute power at the reference area.
Channel Integration Bandwidth
Spectrum Peak Ref n=5 1st element
Spectrum peak power reference at the reference area
Start (Hz) Start frequency for offset
Stop (Hz) Stop frequency for offset
Meas BW (Hz) Measurement bandwidth for offset
Lower(dBm) Absolute peak power on minimum margin point of the negative
offset
Lower lim (dB) Minimum margin from limit line which is decided by Fail Mask
setting on the negative offset
Lower Freq (Hz) Frequency on minimum margin point of the negative offset
Upper (dBm) Absolute peak power on minimum margin point of the positive
offset
Upper lim (dB) Minimum margin from limit line which is decided by Fail Mask
setting on the positive offset
Upper Freq (Hz) Frequency on minimum margin point of the positive offset
Key Path: View/Display
Initial S/W Revision: Prior to A.02.00