755
Modulation Accuracy (Composite Rho)
Meas Setup
Chip Rate
Sets the chip rate.
Multi Channel Estimator
Sets the Multi Channel Estimator On or Off.
IF Gain
Enables you to control an internally switched IF amplifier with approximately 10 dB of gain. This
amplifier takes full advantage of the RF dynamic range of the test set. When it can be turned on without
an overload, the dynamic range is always better when the amplifier is set to On, than when it is set to Off.
The IF Gain key can be used to set the IF Gain function to Auto, or to On (the extra 10 dB) or Off. These
Mode cdma2000
Remote Command [:SENSe]:RHO:CRATe <freq>
[:SENSe]:RHO:CRATe?
Example RHO:CRAT 1228800
RHO:CRAT?
Preset 1.2288 MHz
State Saved Saved in instrument state.
Min 1.105925 MHz
Max 1.351680 MHz
Key Path Meas Setup, Advanced
Mode cdma2000
Remote Command [:SENSe]:RHO:MCEStimator OFF|ON|0|1
[:SENSe]:RHO:MCEStimator?
Example RHO: MCES: ON
RHO: MCES?
Dependencies/Couplings Only if Multi Channel Estimator is On, the view of Power, Timing and Phase
can work. If Multi Channel Estimator is Off, there will be a message “Press
Meas Setup, Advanced and turn Multi Channel Estimation On to see the
data” appears on the view of Power, Timing and Phase.
Preset Off
State Saved Saved in instrument state.
Range On|Off
Key Path Meas Setup, Advanced