655
Code Domain
12 Code Domain Error:
Returns a series of floating point numbers (in dB or dBm) that represents all the code domain
errors.
In BTS mode, there are 64 or 128 numbers depending on Base Code Length.
1st number = 1st code domain error over one PCG specified by Meas Offset
2nd number = 2nd code domain error over one PCG specified by Meas Offset…
N th number = N th code domain error over one PCG specified by Meas Offset
In MS mode, there are 32 I/Q pairs.
1st number = 1st in-phase code domain error over one PCG specified by Meas Offset
2nd number = 1st quad-phase code domain error over one PCG specified by Meas Offset.
(2×N ñ1) number = N th in-phase code domain error over one PCG specified by Meas Offset.
(2×N) number = N th quad-phase code domain error over one PCG specified by Meas Offset.