Modulation Accuracy (Composite Rho)
Meas Setup
Limits
Accesses a menu that allows you to set the following limits:
RMS EVM (Composite)
Peak EVM (Composite)
Rho (Composite)
Peak Code Domain Error
Timing Error
Phase Error
RMS EVM (Composite)
Sets the limit for the composite RMS EVM measurement pass/fail test.
KEYExponential averaging
SCPIEXPonential
When Measure is set at Cont, data acquisitions will continue
indefinitely. After N averages, exponential averaging is used with a
weighting factor of N (the displayed average count stops at N).
Exponential averaging weights new data more than old data, which
allows tracking of slow-changing signals.
KEYRepeat averaging
SCPIREPeat
When Measure is set at Cont, data acquisitions will continue
indefinitely. After N averages is reached, all previous result data is
cleared and the average count is set back to 1. This is equivalent to
being in Measure Single and pressing the Restart key when the
Single measurement finishes.
Mode cdma2000
Remote Command [:SENSe]:RHO:AVERage:TCONtrol EXPonential|REPeat
[:SENSe]:RHO:AVERage:TCONtrol?
Example RHO:AVER:TCON EXP
RHO:AVER:TCON?
Preset REPeat
State Saved Saved in instrument state.
Range Exp|Repeat
Key Path Meas Setup
Key Path Meas Setup
Mode cdma2000