591
Spectrum Emission Mask Measurement
View/Display
Trace Window
Results Window
Limit Lines

Toggles the limit lines display function for the spectrum emission mask measurements On and Off.

Corresponding Trace yellow - Combined trace from carrier and each offset
Name Corresponding Results
Total Pwr Absolute power at the reference area.
Channel Integration Bandwidth
Spectrum Peak Ref n=5 1st element
Peak power at the reference area
Start (Hz) Start frequency for offset
Stop (Hz) Stop frequency for offset
Meas BW (Hz) Measurement bandwidth for offset
Lower Peak (dB) Relative peak power on minimum margin point of the negative
offset
Lower lim (dB) Minimum margin from limit line which is decided by Fail Mask
setting on the negative offset
Lower Peak (dBm) Absolute peak power on minimum margin point of the negative
offset
Upper Peak (dB) Relative peak power on minimum margin point of the positive
offset
Upper lim (dB) Minimum margin from limit line which is decided by Fail Mask
setting on the positive offset
Upper Peak (dBm) Absolute peak power on minimum margin point of the positive
offset
Key Path: View/Display
Initial S/W Revision: Prior to A.02.00
Key Path: View/Display
Mode: WCDMA, C2K, WIMAX OFDMA, TD-SCDMA, 1xEVDO, LTE, LTETDD,
WLAN
Remote Command: :CALCulate:SEMask:LLINe:STATe ON|OFF|1|0
:CALCulate:SEMask:LLINe:STATe?