QPSK EVM Measurement
SPAN X Scale
Ref Value (X Scale, Phase Error Window)Sets the chip reference value on the horizontal axis in the phase error window.Ref Value (X Scale, EVM Window)Sets the chip reference value on the horizontal axis in the EVM window.
Instrument S/W Revision Prior to A.02.00
Remote Command :DISPlay:EVMQpsk:VIEW2:WINDow2:TRACe:X[:SCALe]:RLEVel
<real>
:DISPlay:EVMQpsk:VIEW2:WINDow2:TRACe:X[:SCALe]:RLEVel?
Example DISP:EVMQ:VIEW2:WIND2:TRAC:X:RLEV 1001
DISP:EVMQ:VIEW2:WIND2:TRAC:X:RLEV?
Key Path SPAN X Scale
Mode WCDMA, C2K, 1xEVDO
Notes This key is for control of the reference value of the X scale of the focused
window of the selected view.
Couplings When this parameter has been set, XScaleAutoMag turns off.
Preset 0
State Saved Saved in instrument state.
Min –5000000
Max 5000000
Instrument S/W Revision Prior to A.02.00
Remote Command :DISPlay:EVMQpsk:VIEW2:WINDow3:TRACe:X[:SCALe]:RLEVel
<real>
:DISPlay:EVMQpsk:VIEW2:WINDow3:TRACe:X[:SCALe]:RLEVel?
Example DISP:EVMQ:VIEW2:WIND3:TRAC:X:RLEV 1001
DISP:EVMQ:VIEW2:WIND3:TRAC:X:RLEV?
Key Path SPAN X Scale
Mode WCDMA, C2K, 1xEVDO
Notes This key is for control of the reference value of the X scale of the focused
window of the selected view.
The mode must be in W-CDMA, 1xEVDO or cdma2000 to use this function.
Use INSTrument:SELect to set this mode.
Couplings When this parameter has been set, XScaleAutoEvm turns off.