1378
Trigge r
Input I
Pressing this key, when it is not selected, selects the I channel's ADC voltage as the trigger. The Input I
trigger condition is met when the voltage crosses the trigger level.
Trigger Level
Sets a level for the Input I trigger. When the signal crosses this level, with the chosen slope, the trigger
occurs.
Trig Slop e
Controls the trigger polarity. It is set positive to trigger on a rising edge and negative to trigger on a
falling edge.
Preset: 1 us
OFF
State Saved: Saved in instrument state
Range: –2.5 s to +10 s
Initial S/W Revision: Prior to A.02.00
Key Path: Trigger, Baseband I/Q
Example: TRIG:<meas>:SOUR IINP
Readback Text: Input I
Initial S/W Revision: Prior to A.02.00
Key Path: Trigger, Baseband I/Q, Input I
Remote Command: :TRIGger[:SEQuence]:IINPut:LEVel <voltage>
:TRIGger[:SEQuence]:IINPut:LEVel?
Example: TRIG:IINP:LEV 0.5 V
Preset: 0.25 V
State Saved: Saved in instrument state
Range: –1 to 1 V
Readback Text: 0.1 of displayed unit (V, mV, etc.)
Initial S/W Revision: Prior to A.02.00
Key Path: Trigger, Baseband I/Q, Input I
Remote Command: :TRIGger[:SEQuence]:IINPut:SLOPe POSitive | NEGative
:TRIGger[:SEQuence]:IINPut:SLOPe?
Example: TRIG:IINP:SLOP POS