Sweep/Control
Gate Length Controls the length of time that the gate is on after it opens. MethodThis lets you choose one of the three different types of gating.
Preset: 57.7 us
WiMAX OFDMA: 71 us
GSM/EDGE: 600 us
State Saved: Saved in instrument state
Min: 0.0 us
Max: 100 s
Backwards Compatibility SCPI: [:SENSe]:SWEep:TIME:GATE:DELay ESA compatibility
Initial S/W Revision: Prior to A.02.00
Key Path: Sweep/Control, Gate
Remote Command: [:SENSe]:SWEep:EGATe:LENGth <time>
[:SENSe]:SWEep:EGATe:LENGth?
Example: SWE:EGAT:LENG 1
SWE:EGAT:LENG?
Notes: Units of time are required or no units; otherwise an invalid suffix error
message will be generated.
Dependencies: Grayed out when Gate Method is set to FFT in which case the label changes to
that shown below.
The key is also grayed out if Gate Control = Level.
Preset: 461.6 us
WiMAX OFDMA: 50 us
GSM/EDGE: 200 us
State Saved: Saved in instrument state
Min: 100 ns
Max: 5 s
Backwards Compatibility SCPI: [:SENSe]:SWEep:TIME:GATE:LENGth ESA compatibility
Initial S/W Revision: Prior to A.02.00
Gate Length
(=1.83/RBW)