806
QPSK EVM Measurement
Meas Setup
Meas Interval

Sets the length of the measurement interval (number of data points) that are used.

Remote Command [:SENSe]:EVMQpsk:AVERage:TCONtrol EXPonential|REPeat
[:SENSe]:EVMQpsk:AVERage:TCONtrol?
Example EVMQ:AVER:TCON REP
EVMQ:AVER:TCON?
Key Path Meas Setup
Mode WCDMA, C2K, 1xEVDO
Notes Selects the type of termination control used for averaging. This determines the
averaging action after the specified number of frames (average count) is
reached.
Exponential - Each successive data acquisition after the average count is
reached, is exponentially weighted and combined with the existing average.
Repeat - After reaching the average count, the averaging is reset and a new
average is started.
Preset REPeat
State Saved Saved in instrument state.
Range Exp|Repeat
Instrument S/W Revision Prior to A.02.00
Remote Command [:SENSe]:EVMQpsk:SWEep:POINts <integer>
[:SENSe]:EVMQpsk:SWEep:POINts?
Example EVMQ:SWE:POIN 1001
EVMQ:SWE:POIN?
Key Path Meas Setup
Mode WCDMA, C2K
Preset WCDMA: 2560
C2K: 512
State Saved Saved in instrument state.
Min 128
Max WCDMA: 5120
C2K: 1536
Instrument S/W Revision Prior to A.02.00