691
Code Domain
Meas Setup
quadrature component in demodulation. The correct setting (Normal or Invert) depends on whether the
signal at the input of the instrument has a high or low side mix.
Meas Preset
Restores all the measurement parameters to their default values.
Advanced
Accesses a menu of functions that enable you to set up more specific parameters for the measurement.
Active Set Threshold
Toggles the active channel identification function between Auto and Man. If set to Auto, the active
channels are determined automatically by the internal algorithm. If set to Man, the active channel
identification is determined by a user definable threshold ranging from 0.00 to –100.00 dB.
Mode cdma2000
Remote Command [:SENSe]:CDPower:SPECtrum INVert|NORMal
[:SENSe]:CDPower:SPECtrum?
Example CDP:SPEC INV
CDP:SPEC?
Preset NORMal
State Saved Saved in instrument state.
Range Normal|Invert
Key Path Meas Setup
Mode cdma2000
Remote Command :CONFigure:CDPower
Example CONF:CDP
Dependencies/Couplings Selecting Restore Measurement Defaults restores all measurement
parameters to their default values.
Key Path Meas Setup
Key Path Meas Setup
Mode cdma2000