Channel Power Measurement
Meas Setup
Avg Mode
Allows you to select the type of termination control used for the averaging function. This determines the
averaging action after the specified number of data acquisitions (average count) is reached.
When set to Exponential (Exp) the measurement averaging continues using the specified number of
averages to compute each exponentially-weighted averaged value. The average is displayed at the end of
each sweep.
When set to Repeat, the measurement resets the average counter each time the specified number of
averages is reached.
Preset: WCDMA: 200
WIMAX OFDMA, LTE, LTETDD: 200
CDMA2K: 20
1xEVDO: 20
WLAN: 10
ON
State Saved: Saved in instrument state.
Min: 1
Max: 10000
Initial S/W Revision: Prior to A.02.00
Modified at S/W Revision: A.02.00, A.03.00
Key Path: Meas Setup
Mode: WCDMA, C2K, WIMAX OFDMA, 1xEVDO, LTE, LTETDD, WLAN
Remote Command: [:SENSe]:CHPower:AVERage:TCONtrol EXPonential|REPeat
[:SENSe]:CHPower:AVERage:TCONtrol?
Example: CHP:AVER:TCON EXP
CHP:AVER:TCON?
Notes: You must be in the W-CDMA mode, cdma2000 mode, LTD mode, LTE TDD
mode, WLAN mode, 1xEVDO mode, or WIMAX OFDMA mode to use this
command. Use :INSTrument:SELect to set the mode.
Preset: EXP
State Saved: Saved in instrument state.
Range: Exp|Repeat
Initial S/W Revision: Prior to A.02.00
Modified at S/W Revision: A.02.00, A.03.00