627
Occupied Bandwidth Measurement
Save
Save
See “Save” on page 204 for more information.
Data
See “Data (Export)” on page 213 for more information.
Measurement Results
Pressing this key selects Meas Results as the data type to be exported.
The Meas Results file contains measurement result sets, plus information describing the current state of
the analyzer, as detailed in “Meas Results File Definition” on page 627 and “Meas Results File
Example” on page 629 below.
Meas Results File Definition
The content of a Meas Results File is defined in this section.
The first lines in the file consist of identification and instrument configuration details, as follows.
Key Path: Front-panel key
Initial S/W Revision: Prior to A.02.00
Key Path: Front-panel key
Key Path: Save, Data
Remote Command: :MMEMory:STORe:RESults <string>
Example: :MMEM:STOR:RES “MeasR_0000.csv”
Notes: If the save is initiated via SCPI and the file already exists, the file will be
overwritten.
The SCPI command exports Occupied Bandwidth measurement results to the
file specified as the parameter in the current path. The default path is My
Documents\<current mode>\data\OBW\results.
Using the C: drive is strongly discouraged, since it runs the risk of being
overwritten during an instrument software upgrade.
The SCPI parameter is a quoted string that specifies the filename. Both single
and double quotes are supported for any filename parameter over SCPI.
Dependencies: The current active measurement must be the Occupied Bandwidth
measurement to use this command.
Status Bits/OPC dependencies: Sequential – waits for the previous measurement to complete
Initial S/W Revision: Prior to A.02.00