1309
Trace/Detector
Trace Update State On/Off
Trace Display State On/Off
More Information
When a trace becomes inactive, the following things happen:
Any update from the SENSe system (detectors) immediately stops (does not wait for end of sweep)
the trace is displayed at half intensity (as long as it stays inactive)
Inactive traces display across the entire X-axis of the instrument. Their horizontal placement does not
change even if X-axis settings subsequently are changed, although Y-axis settings will affect the vertical
placement of data.
In most cases, inactive traces are static and unchanging; however, there are cases when an inactive trace
will update, specifically:
if data is written to that trace from remote
if trace data is loaded from mass storage
Key Path Trace/Detector
Remote Command :TRACe[1]|2|3|4|5|6:UPDate[:STATe] ON|OFF|0|1
:TRACe[1]|2|3|4|5|6:UPDate[:STATe]?
Example TRAC2:UPD 0 Makes trace 2 inactive (stops updating)
Couplings Whenever you set Update to On for any trace, the Display is set to On for that
trace.
Preset 1|0|0|0|0|0 (On for Trace 1; Off for 2–6)
State Saved Saved in Instrument State
Initial S/W Revision Prior to A.02.00
Key Path Trace/Detector
Remote Command :TRACe[1]|2|3|4|5|6:DISPlay[:STATe] ON|OFF|0|1
:TRACe[1]|2|3|4|5|6:DISPlay[:STATe]?
Example TRAC2:DISP,1 Makes trace 2 visible
TRAC3:DISP,0 Blanks trace 3
Couplings Whenever you set Update to On for any trace, the Display is set to On for that
trace.
Preset 1|0|0|0|0|0 (On for Trace 1; Off for 2–6)
State Saved Saved in Instrument State
Initial S/W Revision Prior to A.02.00