730
Modulation Accuracy (Composite Rho)
AMPTD Y Scale
Attenuation
Accesses a menu of functions that enable you to change the attenuation settings. This key has read-back
text that describes the total attenuator value.
See “Attenuation” on page 850 for more information.
Scale/Div
Sets the sensitivity for the vertical axis. The range is 0.10 to 50.00 dB with 0.01 dB resolution. However,
since Auto Scaling defaults to On, this value is automatically determined by the measurement result.
When you set a value manually, Auto Scaling automatically changes to Off.
I/Q Error View - EVM Window
Sets the sensitivity measurement result in the EVM window.
Min –36000.0
Max 36000.0
Key Path AMPTD Y Scale
Key Path AMPTD Y Scale
Key Path AMPTD Y Scale
Mode cdma2000
Remote Command :DISPlay:RHO:VIEW2:WINDow[1]:TRACe:Y[:SCALe]:PDIVi
sion <real>
:DISPlay:RHO:VIEW2:WINDow[1]:TRACe:Y[:SCALe]:PDIVi
sion?
Example DISP:RHO:VIEW2:WIND:TRAC:Y:PDIV 10.0
DISP:RHO:VIEW2:WIND:TRAC:Y:PDIV?
Restriction and Notes If Auto Scaling is set to On, this value is automatically determined by
the measurement result. When you set this value manually, Auto
Scaling is automatically set to Off.
Dependencies/Couplings See Restriction and Notes
Preset 0.5
State Saved Saved in instrument state.
Min 0.100
Max 50.0
Key Path AMPTD Y Scale