1381
Trigger
Auxiliary Channel I/Q Mag
Pressing this key, when it is not selected, selects the Auxiliary Channel I/Q magnitude signal as the
trigger. The Auxiliary Channel I/Q Magnitude trigger condition is met when the auxiliary receiver's I/Q
magnitude output crosses the Auxiliary I/Q magnitude trigger level.
Trigger Level
Sets a level for the I/Q magnitude trigger. When the signal crosses this level, with the chosen slope, the
trigger occurs.
Trig Slo pe
Controls the trigger polarity. It is set positive to trigger on a rising edge and negative to trigger on a
falling edge.
Key Path: Trigger, Baseband I/Q
Example: TRIG:<meas>:SOUR AIQM
Readback Text: Aux I/Q Mag
Initial S/W Revision: Prior to A.02.00
Key Path: Trigger, Baseband I/Q, Aux Channel I/Q Mag
Remote Command: :TRIGger[:SEQuence]:AIQMag:LEVel <ampl >
:TRIGger[:SEQuence]:AIQMag:LEVel?
Example: TRIG:AIQM:LEV –30 dBm
Notes: The I/Q reference impedance is used for converting between power and
voltage.
Preset: –25 dBm
State Saved: Saved in instrument state
Range: –200 dBm to 100 dBm
Readback Text: <level> dBm
Initial S/W Revision: Prior to A.02.00
Key Path: Trigger, Baseband I/Q, Aux Channel I/Q Mag
Remote Command: :TRIGger[:SEQuence]:AIQMag:SLOPe POSitive | NEGative
:TRIGger[:SEQuence]:AIQMag:SLOPe?
Example: TRIG:AIQM:SLOP POS
Preset: POSitive
State Saved: Saved in instrument state
Initial S/W Revision: Prior to A.02.00