352
Channel Power Measurement
Save
Save
See “Save” on page 204 for more information.
Data
See “Data (Export)” on page 213 for more information.
Measurement Results
Pressing this key selects Meas Results as the data type to be exported.
The Meas Results file contains information which describes the current state of the analyzer. It is
detailed in Meas Result File Contents below.
Meas Results File Contents
A Meas Results File contains measurement results with the following information.
File ID string, which is “MeasResult”
Measurement ID following Mode ID, which is “SA:CHP” for example.
Firmware rev and model number
Key Path: Front-panel key
Key Path: Front-panel key
Key Path: Save, Data
Remote Command: :MMEMory:STORe:RESults <string>
Example: :MMEM:STOR:RES “MeasR_0000.csv”
Notes: If the save is initiated via SCPI and the file already exists, the file will be
overwritten.
The SCPI command exports Channel Power measurement results to the file
specified as the parameter in the current path. The default path is My
Documents\<current mode>\data\CHP\results.
Using the C: drive is strongly discouraged, since it runs the risk of being
overwritten during an instrument software upgrade.
The SCPI parameter is a quoted string, which specifies the filename. Both
single and double quotes are supported for any filename parameter over SCPI.
Dependencies: The current active measurement must be the Channel Power measurement to
use this command.
Status Bits/OPC dependencies: Sequential – waits for the previous measurement to complete.
Initial S/W Revision: Prior to A.02.00