1047
Meas Setup
Meas Setup
The Meas Setup key opens up a menu of keys that allow you to control the most important parameters
for the current measurement.
NOTE In the Meas Setup menu you may configure Averaging, by setting the Average
Number and the Average Type.
Average/Hold Number
Sets the terminal count number N for Average, Max Hold and Min Hold trace types. This number is an
integral part of how the average trace is calculated. Basically, increasing N results in a smoother average
trace.
See “More Information” on page 1047.
See “AVER:CLE command” on page 1047.
More Information
AVER:CLE command
The AVER:CLE command (below) resets the average/hold count and does an INIT:IMM, which begins
Key Path Front-panel key
Initial S/W Revision Prior to A.02.00
Key Path Meas Setup
Remote Command [:SENSe]:AVERage:COUNt <integer>
[:SENSe]:AVERage:COUNt?
Couplings Restarting any of these functions (Average, Max Hold or Min Hold) restarts
all of them, as there is only one count.
Preset 100
State Saved Saved in Instrument State
Min 1
Max 10000
Status Bits/OPC dependencies See the Section “Sweep/Control” on page 1267 for a discussion of the
Sweeping, Measuring, Settling and OPC bits, and the Hi Sweep line. All are
affected when a sequence is reset.
Initial S/W Revision Prior to A.02.00