1210
Source (Internal)
Marker 1
Sets marker 1 to be used for the pulse/RF blanking function.
Marker 2
Sets marker 2 to be used for the pulse/RF blanking function.
Marker 3
Sets marker 3 to be used for the pulse/RF blanking function.
Marker 4
Sets marker 4 to be used for the pulse/RF blanking function.
ALC Hold
Allows you to specify which marker is routed for use within the ALC hold function. The ALC hold marker
function holds the ALC circuitry at the average value of the sample points set by the marker.
The ALC hold function operates during the low periods of the marker signal. The marker polarity determines when
the marker signal is high. For positive polarity, this is during the marker points. For a negative polarity, this is when
there are no maker points.
Key Path: Source, Modulation Setup, ARB, Marker Utilities, Marker Routing,
Pulse/RF Blank
Example: SOUR:RAD:ARB:MDES:PULS M1
Initial S/W Revision: A.05.00
Key Path: Source, Modulation Setup, ARB, Marker Utilities, Marker Routing,
Pulse/RF Blank
Example: SOUR:RAD:ARB:MDES:PULS M2
Initial S/W Revision: A.05.00
Key Path: Source, Modulation Setup, ARB, Marker Utilities, Marker Routing,
Pulse/RF Blank
Example: SOUR:RAD:ARB:MDES:PULS M3
Initial S/W Revision: A.05.00
Key Path: Source, Modulation Setup, ARB, Marker Utilities, Marker Routing,
Pulse/RF Blank
Example: SOUR:RAD:ARB:MDES:PULS M4
Initial S/W Revision: A.05.00
Key Path: Source, Modulation Setup, ARB, Marker Utilities, Marker Routing