1135
Single (Single Measurement/Sweep)
Single (Single Measurement/Sweep)

Sets the test set for Single measurement operation. The single/continuous state is Meas Global, so the

setting will affect all the measurements. If you are Paused, pressing Single does a Resume.

Key Path: Front-panel key
Example: :INIT:CONT OFF
Notes: See Cont key description.
Backwards Compatibility Notes: For Spectrum Analysis mode in ESA and PSA, the Single hardkey and the
INITiate:IMM switched from continuous measurement to single measurement
and restarted sweeps and averages (displayed average count reset to 1), but
did not restart Max Hold and Min Hold. In the X-Series, the Single hardkey
and the INITiate:IMM command initiate a sweep/ measurement/ average
sequence/hold sequence including Max Hold and Min Hold.
For Spectrum Analysis mode in ESA and PSA, the Single hardkey restarted
the sweep regardless of whether or not you were in an active sweep or sweep
sequence. In the X-Series, Restart does this but Single only restarts the
sweep or sweep sequence if you are in the idle state.
Initial S/W Revision: Prior to A.02.00