Modulation Accuracy (Composite Rho)
SPAN X Scale
I/Q Error View - Phase Error Window
Sets the X axis reference position in the phase error window.
Auto Scaling
Toggles the Auto Scaling function between On and Off. When Auto Scaling is On, and the Restart
front-panel key is pressed, this function automatically determines the scale per division and reference
values based on the measurement results.
When you set a value to either Scale/Div or Ref Value manually, Auto Scaling automatically changes to
Off.
I/Q Error View - EVM Window
When Auto Scaling is On and the Restart front-panel key is pressed, will result in automatically
displaying the scale per division and reference value results in the EVM window.
Key Path Span X Scale
Mode cdma2000
Remote Command :DISPlay:RHO:VIEW2:WINDow3:TRACe:X[:SCALe]:RPOSiti
on LEFT|CENTer|RIGHt
:DISPlay:RHO:VIEW2:WINDow3:TRACe:X[:SCALe]:RPOSiti
on?
Example DISP:RHO:VIEW2:WIND3:TRAC:X:RPOS LEFT
DISP:RHO:VIEW2:WIND3:TRAC:X:RPOS?
Preset LEFT
State Saved Saved in instrument state.
Range Left|Ctr|Right
Key Path Span X Scale
Key Path Span X Scale
Mode cdma2000
Remote Command :DISPlay:RHO:VIEW2:WINDow[1]:TRACe:X[:SCALe]:COUPl
e 0|1|OFF|ON
:DISPlay:RHO:VIEW2:WINDow[1]:TRACe:X[:SCALe]:COUPl
e?
Example DISP:RHO:VIEW2:WIND:TRAC:X:COUP ON
DISP:RHO:VIEW2:WIND:TRAC:X:COUP?