409
ACP Measurement
Meas Setup
Reference” value.PSDSets the power spectral density in the carrier (main channel) that is used to compute the relative power spectral density values for the offsets when Meas Type is set to PSD Ref. When the PSD Ref state is set to Auto, this will be set to the measured carrier power spectral density.Key Path Meas Setup, Carrier Setup, Power Ref
Mode WCDMA, C2K, WIMAX OFDMA, TD-SCDMA, LTE, LTETDD
Remote Command [:SENSe]:ACPower:CARRier[1]|2[:POWer] <real>
[:SENSe]:ACPower:CARRier[1]|2[:POWer]?
[:SENSe]:ACPower:CARRier[1]|2:AUTO[:STATe] OFF|ON|0|1
[:SENSe]:ACPower:CARRier[1]|2:AUTO[:STATe]?
Example ACP:CARR 10
ACP:CARR?
ACP:CARR:AUTO OFF
ACP:CARR:AUTO?
Notes Although the default value is defined, the value is recalculated by the
measurement result just after measurement.
Carrier sub op code. 1 for BTS, 2 for MS. Default is BTS.
You must be in the mode that includes ACP measurements to use this
command. Use :INSTrument:SELect to set the mode.
Dependencies This key is available only when the Meas Type is TPRef. If the Meas Type is
not TPRef, this key is grayed out.
Preset 0.0
ON
State Saved Saved in instrument state.
Min –200 dBm
Max 200 dBm
Backwards Compatibility SCPI [:SENSe]:MCPower:CARRier[1]|2[:POWer]
Initial S/W Revision Prior to A.02.00
Modified at S/W Revision A.02.00, A.03.00, A.04.00
Key Path Meas Setup, Carrier Setup, Power Ref
Mode WCDMA, C2K, WIMAX OFDMA, TD-SCDMA, LTE, LTETDD