1274
Sweep/Control
whether you want to optimize for dynamic range or for speed. These rules are chosen under the Sweep
Type R ules key.
Swept Manually selects swept analysis, so it cannot change automatically to FFT. FFT Manually selects FFT analysis, so it cannot change automatically to Swept.
Key Path: Sweep/Control, Sweep Setup, Sweep Type
Remote Command: [:SENSe]:SWEep:TYPE:AUTO OFF|ON|0|1
[:SENSe]:SWEep:TYPE:AUTO?
Example: :SWE:TYPE:AUTO ON
Couplings: Pressing Auto Couple always sets Sweep Type to Auto.
Swept is always chosen whenever any form of Signal ID is on, or the Source
Mode is set to Tracking, or any EMI detector is selected, or the RF Preselector
is ON.
Preset: ON
State Saved: Saved in instrument state
Initial S/W Revision: Prior to A.02.00
Key Path: Sweep/Control, Sweep Setup, Sweep Type
Example: SWE:TYPE SWE
Dependencies: Grayed out while in Gated FFT (meaning Gate is ON and Gate Method is
FFT).
If this key is selected, the gate method Gated FFT is grayed out.
Couplings: This selection is chosen automatically if any of the CISPR detectors is chosen
for any active trace, in which case the FFT Sweep Type selection is also
grayed out.
State Saved: Saved in instrument state
Readback: Swept
Initial S/W Revision: Prior to A.02.00
Modified at S/W Revision: A.02.00
Key Path: Sweep/Control, Sweep Setup, Sweep Type
Example: SWE:TYPE FFT