Spectrum Emission Mask Measurement
Trace/Detector
Trace/Detector

Accesses a menu of functions that enable you to control trace and detector for the current measurement.

Trac e Typ e

Allows you to select the type of trace for the current measurement. The menu contains a 1–of–N

selection of the trace type (Clear Write, Average, Max Hold, Min Hold).

Chan Detector

Accesses a menu of functions that enable you to control the detectors for reference channel. The

following choices are available:

Key Path: Front-panel key
Initial S/W Revision: Prior to A.02.00
Key Path: Trace/Detector
Mode: WCDMA, C2K, WIMAX OFDMA, TD-SCDMA, 1xEVDO, LTE, LTETDD,
WLAN
Remote Command: :TRACe:SEMask:TYPE WRITe|AVERage|MAXHold|MINHold
:TRACe:SEMask:TYPE?
Example: TRAC:SEM:TYPE MINH
TRAC:SEM:TYPE?
Notes: WRITe = Clear Write
AVERage = Average
MAXHold = Maximum Hold
MINHold = Minimum Hold
Couplings: When Detector setting is “Auto” ([:SENSe]:SEMask:DETector:AUTO?),
Detector ([:SENSe]:SEMask:DETector[:FUNCtion]?) switches aligning with
the switch of this parameter: “NORMal” with WRITe (Clear Write),
“AVERage” with AVERage, “POSitive (peak)” with MAXHold, and
“NEGative (peak)” with MINHold.
Preset: AVERage
State Saved: Saved in instrument state.
Range: WRITe|AVERage|MAXHold|MINHold
Initial S/W Revision: Prior to A.02.00
Modified at S/W Revision: A.02.00, A.03.00