382
ACP Measurement
AMPTD Y Scale
AMPTD Y Scale
Accesses a menu of functions that enable you to set the vertical scale parameters. The parameter values
are measurement independent, except all Attenuation values and the Internal Preamp selections, which
are the same across all measurements.
Ref Value
Sets the value for the absolute power reference. However, since the Auto Scaling is defaulted to On, this
value is automatically determined by the measurement result. When you set a value manually, Auto
Scaling automatically changes to Off.
Attenuation
Accesses a menu of functions that enable you to change attenuation settings. This key has read-back text
that describes the total attenuator value.
Key Path Front-panel key
Initial S/W Revision Prior to A.02.00
Key Path AMPTD Y Scale
Mode WCDMA, C2K, WIMAX OFDMA, TD-SCDMA, 1xEVDO, LTE, LTETDD
Remote Command :DISPlay:ACPower:VIEW[1]:WINDow[1]:TRACe:Y[:SCALe]:R LEV
el <real>
:DISPlay:ACPower:VIEW[1]:WINDow[1]:TRACe:Y[:SCALe]:RLEV
el?
Example DISP:ACP:VIEW:WIND:TRAC:Y:RLEV 100
DISP:ACP:VIEW:WIND:TRAC:Y:RLEV?
Notes You must be in the mode that includes ACP measurements to use this
command. Use :INSTrument:SELect to set the mode.
Couplings When the Auto Scaling is On, this value is automatically determined by the
measurement result.
When you set a value manually, Auto Scaling automatically changes to Off.
Preset –30 dBm
State Saved Saved in instrument state.
Min –250.00 dBm
Max 250.00 dBm
Initial S/W Revision Prior to A.02.00
Modified at S/W Revision A.02.00, A.03.00