1358
Trigge r
Trig Slop e Controls the trigger polarity. It is set positive to trigger on a rising edge and negative to trigger on a
falling edge.
Trig Delay Controls a time delay during which the analyzer will wait to begin a sweep after meeting the trigger criteria. You can use negative delay to pre-trigger the instrument in time domain or FFT, but not in swept
Remote Command: :TRIGger[:SEQuence]:EXTernal2:LEVel
:TRIGger[:SEQuence]:EXTernal2:LEVel?
Example: TRIG:EXT2:LEV 1.1 V
Couplings: This same level is used for the Ext2 trigger source in the Trigger menu, for the
Ext2 selection in the Periodic Timer sync source (in the Trigger menu and in
the Gate Source menu), and also for the Ext2 selection in the Gate Source
menu.
Preset: 1.2 V
State Saved: Saved in instrument state
Min: –5 V
Max: 5 V
Default Unit: V
Backwards Compatibility SCPI: :TRIGger[:SEQuence]:FRAMe:EXTernal2:LEVel
Initial S/W Revision: Prior to A.02.00
Key Path: Trigger, External 2
Remote Command: :TRIGger[:SEQuence]:EXTernal2:SLOPe POSitive|NEGative
:TRIGger[:SEQuence]:EXTernal2:SLOPe?
Example: TRIG:EXT2:SLOP NEG
Couplings: This same slope is used in the Ext2 selection for the trigger source in the
Trigger menu and for the period timer sync source (in the Trigger menu and in
the Gate Source menu).
Preset: POSitive
State Saved: Saved in instrument state
Backwards Compatibility SCPI: :TRIGger[:SEQuence]:FRAMe:EXTernal2:SLOPe
Backwards Compatibility Notes: The legacy :TRIGger[:SEQuence]:SLOPe command affects the slopes for the
VID, LINE, EXT1, EXT2, and RFB triggers.
Initial S/W Revision: Prior to A.02.00