225
System Functions
System
Refresh
When pressed, refreshes the Show Errors display.
Clear Error Queue
This clears all errors in all error queues.
Note the following:
Clear Error Queue does not affect the current status conditions.
Mode Preset does not clear the error queue.
Restore System Defaults will clear all error queues.
*CLS only clears the queue if it is sent remotely and *RST does not affect any error queue.
Switching modes does not affect any error queues.
System
The System screen is formatted into three groupings: product descriptive information, options tied to the
hardware, and software products:
Initial S/W Revision: A.10.01
Key Path: System, Show, Errors
Initial S/W Revision: Prior to A.02.00
Key Path: System, Show, Errors
Initial S/W Revision: Prior to A.02.00
<Product Name> <Product Description>
Product Number: N9020A
Serial Number: US46220924
Firmware Revision: A.01.01
Computer Name: <hostname>
Host ID: N9020A,US44220924
N9020A-503 Frequency Range to 3.6 GHz
N9020A-PFR Precison Frequency Reference
N9020A-P03 Preamp 3.6 GHz
N9060A-2FP Spectrum Analysis Measurement Suite 1.0.0.0
N9073A-1FP WCDMA 1.0.0.0
N9073A-2FP WCDMA with HSDPA 1.0.0.0