Input/Output
More Information
Multiport adapter in the EXT (E6607A/B) supports the Agilent E6617A, which provides a USB
connection for download of calibration data and additional control.
The connection diagram for Agilent E6617A switch unit is:
Multiport Adapter On/Off
Turning the Multiport Adapter On means that the multiport adapter unit is connected and it will be used
for the measurements or source.
Turning the Multiport Adapter Off means that the multiport adapter unit will not be used for the analyzer
Dependencies: Multiport adapter is only available for EXT (E6607A/B) model, and is
blanked for other models.
Multiport adapter unit is connected to EXT by USB for download of
calibration data and additional control.
Preset: All settings under this key are returned to their default state when Restore
Input/Output Defaults is pressed.
State Saved: All settings under this key, are remembered when you unplugged the multiport
adapter unit USB connection, so that when multiport adapter unit USB is
connected again, all the multiport adapter functions will retain their previous
settings, with the exception of Multiport Adapter which is set to OFF.
Initial S/W Revision: A.10.00
RF Input
RF Output
RF IO
TRIGGER 1
OUT TRIGGER 2
OUT USB Cable