1088
Meas Setup
Off
Pressing this key, turns the demodulation function off.
Demod Time
Sets the amount of time the test set demodulates the signal after each sweep. The demodulated signal can
be heard through the speaker during demodulation. In zero span, demodulation can be performed
continuously, making this parameter not applicable, hence it is grayed out in zero span.
Couplings In zero span only, the value is set equal to the test set’s current RBW value and
it displays that value on the key, but the key is grayed out.
Preset 100 kHz
State Saved Saved in instrument state.
Min 390 Hz
Max 8 MHz
Default Unit Hz
Initial S/W Revision Prior to A.02.00
Key Path Meas Setup, Analog Demod Tune&Listen
Example DEM OFF turns the demodulation function OFF
State Saved Saved in instrument state.
Initial S/W Revision Prior to A.02.00
Key Path Meas Setup, Analog Demod Tune&Listen
Remote Command [:SENSe]:DEMod:TIME <time>
[:SENSe]:DEMod:TIME?
Example DEM:TIME 500 ms
DEM:TIME?
Notes This key / command is grayed out in zero span
Dependencies Unavailable in zero span.
Preset 500 ms
State Saved Saved in instrument state.
Min 2 ms
Max 100 s
Initial S/W Revision Prior to A.02.00