367
Channel Power Measurement
Trace/Detector
Auto
Sets the detector for the currently selected trace to Auto.
Couplings: When Detector setting is “Auto” ([:SENSe]:CHPower:DETector:AUTO?),
Detector ([:SENSe]:CHPower:DETector[:FUNCtion]?) switches aligning
with the switch of this parameter: “NORMal” with Clear Write, “AVERage”
with AVERage, “POSitive (peak)” with MAXHold, and “NEGative (peak)”
with MINHold.
Preset: AVERage
State Saved: Saved in instrument state.
Range: Normal|Average|Peak|Sample|Negative Peak
Initial S/W Revision: Prior to A.02.00
Modified at S/W Revision: A.02.00, A.03.00
Key Path: Trace/Detector
Mode: WCDMA, C2K, WIMAX OFDMA, 1xEVDO, LTE, LTETDD, WLAN
Remote Command: [:SENSe]:CHPower:DETector:AUTO ON|OFF|1|0
[:SENSe]:CHPower:DETector:AUTO?
Example: CHP:DET:AUTO ON
CHP:DET:AUTO?
Couplings: When Detector setting is “Auto” ([:SENSe]:CHPower:DETector:AUTO?),
Detector ([:SENSe]:CHPower:DETector[:FUNCtion]?) switches aligning
with the switch of this parameter: “NORMal” with Clear Write, “AVERage”
with AVERage, “POSitive (peak)” with MAXHold, and “NEGative (peak)”
with MINHold.
Preset: ON
State Saved: Saved in instrument state.
Initial S/W Revision: Prior to A.02.00
Modified at S/W Revision: A.02.00, A.03.00